PULSE
No No 3 The detected pulse signal
is too weak.
C.2.8 IBP Module Alarm Messages
Alarm message A B L Cause Measure
IBP1 SENSOR No Yes 3 The invasive blood pressure Check if the IBP 1 sensor
OFF cable of channel 1 is is properly connected.
disconnected from the monitor.
IBP2 SENSOR No Yes 3 The invasive blood pressure Check if the IBP 2 sensor
OFF cable of channel 2 is is properly connected.
disconnected from the monitor.
IBP1 NEED No No 3 The IBP transducer of channel 1 Zero the IBP transducer
ZERO-CAL has not been zeroed. of channel 1.