Figure 15-6 IBP Alarm Setup Menu
IBP alarm limits
Pressure Label Max. Upper (mmHg) Min. Lower (mmHg) Step (mmHg)
ART 300 0 1
PA 120 -6 1
CVP 40 -10 1
RAP 40 -10 1
LAP 40 -10 1
ICP 40 -10 1
P1 300 -50 1
P2 300 -50 1
The PR alarm limits are the same as those of the corresponding SpO2 module.
SCALE ADJUST >> This option allows you to open the following menu and set
the positions of the high, reference and low scales for the
two waveforms.
Figure 15-7 IBP Scale Adjust Menu
15-6